Using in Series/Parallel Operation
------------------------------------------------------------------------------------
124
Operation
1.Turn on the POWER switches for all power supplies (#1 to #n).
There is no set order to turn on power.
2.After all power is turned on, wait approximately 2 seconds before operating.
3.The ZX power supply which its M/S CONTROL Connector IN is not connected to another ZX-S Series power
supply, is the Master unit (pervious page labeled #1).
4.After the POWER switch has been turned ON, the model (top), and firmware version and number of parallel
connected devices (bottom) will be displayed on the Master unit’s Number Display.
If there is a disagreement with the number of parallel connected devices (there is a difference between the
previous number of parallel connected devices and current number), the error code “E011” will display on the
top Number Display and the detected number of devices will appear blinking on the bottom Number Display.
After verifying the number of parallel devices, hold down the “FUNC.” key for more than 2 seconds to confirm
detected number and return to measurement display mode.
5.Voltage measurement is displayed in the top and current measurement is displayed in the bottom Number Display.
Total current value is displayed in the bottom Number Display.
Select power display with the “Display” key to display total power values.
Error Codes
1.Error Codes & Display List for Master Unit
Place
Items
No. Display
Top
No. Display
Bottom
Content
WDT
E001
-
Malfunction of CPU
AD_OV
E002
-
Input voltage error for DC/DC block
OHP
E003
OHP
Over Heating Protection triggered
OVP
E004
OUP
Over Voltage Protection triggered
OCP
E005
OCP
Over Current Protection triggered
BST
E006
-
Slave error (Displayed only on the Master during
Parallel Operation)
P_ERR
E007
-
Input voltage error
BST_NRDY
E008
-
Slave input voltage error
Device Detection Error
E009
-
No. of parallel devices cannot be detected
Max. Device No. OVER
E010
-
11 or more parallel connected devices
Device No.
Disagreement
E011
-
Difference in the current and previous no. of
connected parallel devices
TRIP(1600LN/LAN)
E012
Shdn
TRIP function triggered
Fan Alarm
E013
FAn
Detected fan alarm
Communication Error
E100
-
E2P Error
E110
-
Initialization error during startup
E2P Error(for LAN)
E120
-
Initialization error during startup
Startup Sequence
Failure
E130
-
Initialization error during startup
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
Slaves, when operating normally, will
display bars as shown on the right.
Содержание ZX-S-LN Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...