Calibration
------------------------------------------------------------------------------------
163
Offset/Full-Scale Calibration of Output Voltage with External Analog Values
(ZX-S Series Type-LAN)
Output Voltage Control with External Voltage
1. Disconnect the load and remote sensing wires and
connect the digital multimeter to the output terminal.
Turn the POWER switch ON and
OUTPUT OFF in advance.
2. Change the top Number Display to Settings Item 70
with the VOLTAGE dial.
3. Change set value to 1 with the CURRENT dial.
1: External Voltage(0~10V)
4. Press FUNC. key again to finalize settings.
5. Press the OVP key ①. Set to “88.00V” with the
VOLTAGE dial ②.
6. Input an external voltage of “0.00V”.
7. Press the OUTPUT key to set output to ON.
Calibration Circuit
Digital
Multimeter
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
CC
CV
OUTPUT POWER (%)
STEP
CURRENT
STEP
VOLTAGE
PRESET FUNC. REMOTE LOCK
20
60
100 LMT
40
80
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
+
-
V
Power Supply
+
-
IN
OUT
4
5
外部電圧
0~10V
External
Voltage
10V
OUTPUT
PRESET
OVP
OCP
FUNC.
ALARM RESET
IN
OUT
4
5
外部電圧
0~10V
External
Voltage
0~10V
OUT PU T
PR ES ET
OVP
OCP
POWE R
FU NC.
LOCAL
LOCK
D ISPL AY
ST ORE
MEMOR Y
CC
CV
OUTPU T POWER
(%)
ST EP
CUR RE NT
ST EP
V OL TAGE
PRE SET FUN C. REMOTE LOCK
A LAR M R ESE T
20
60
100 LMT
40
80
①
②
ZX-S-4 00 LN
A
B
C
Содержание ZX-S-LN Series
Страница 1: ...Constant Voltage Constant Current DC Power Supply series Instruction Manual...
Страница 13: ...About this Product 10 This is blank page...
Страница 63: ...Useful Function 60 This is blank page...
Страница 151: ...Controlling with External Analog Contact Signals 148 This is blank page...
Страница 157: ...Special Loads 154 This is blank page...
Страница 199: ...Specifications 196 This is blank page...
Страница 200: ...197 External Diagrams ZX S 400LN 400LAN 4 M3...
Страница 201: ...Appendix 198 ZX S 800LN 800LAN...
Страница 202: ...Appendix 199 ZX S 1600LN 1600LAN...