CHAPTER 6
Operation
Sysmex XE-5000 Instructions for Use
6-35
R
e
vi
se
d A
p
ri
l 20
07
8. When the ENTER key is pressed, a background check is
run at the same time. Background check analysis is
repeated three times for the background check. If the
background value is lower or equal to that shown in the
table on the left, the background check is completed.
If the background value does not become lower or equal
to the specified value, the message “Background Error”
will be displayed on completion of the background check.
In the case of a “Background Error”, select “Auto Rinse”
on the function menu of the Background Check screen to
perform an automatic cleaning. If the value after the
automatic cleaning is not lower than the acceptable
background value, “+” is displayed to the right of the value.
To resolve background error, see “Chapter 10
Troubleshooting”.
9. If the result of the background check is such that the
background value is lower or equal to the acceptable limit,
the READY LED lights and the Main Unit enters body fluid
analysis ready status.
10. Mix the contents of the test tube gently but thoroughly.
Information
In the case of a “Background Error”, the analysis can
be performed by selecting “OK” on the function
menu. However, the measured values tend to be
higher and correct results may not be obtained.
Information
• The basic analysis procedure is the same as for
normal mode, but for body fluid analysis it is
necessary to reliably eliminate background
influences. That is why the background check is
performed automatically.
• When the system is in body fluid mode, it cannot
perform QC analysis or calibration. Sampler analysis
and manual closed analysis are unavailable.
Analysis orders will not be accepted from a
transportation system (if one is connected).
WBC-BF
0.001 [×10
3
/
µ
L]
RBC-BF
0.003 [×10
6
/
µ
L]
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...