CHAPTER 11
Technical Information
11-4
Sysmex XE-5000 Instructions for Use
Revised July
2
007
Precision
(reproducibility)
Capillary mode
WBC
9.0% or less (4.0 × 10
3
/
µ
L or more)
RBC
4.5% or less (4.00 × 10
6
/
µ
L or more)
HGB
3.0% or less
HCT
4.5% or less
MCV
4.5% or less
MCH
4.5% or less
MCHC
4.5% or less
PLT
12.0% or less (100 × 10
3
/
µ
L or more)
RET#
35% or less (RBC 3.00 × 10
6
/
µ
L or more, RET% 1-4%)
RET%
35% or less (RBC 3.00 × 10
6
/
µ
L or more, RET% 1-4%)
WBC-D
9.0% or less (4.0 × 10
3
/
µ
L or more)
NEUT%
24.0% or less (30.0 NEUT% or more, WBC 4.0 × 10
3
/
µ
Lor more)
LYMPH% 24.0% or less (15.0 LYMPH% or more, WBC 4.0 × 10
3
/
µ
L or more)
MONO% 60.0% or less ( 5.0 MONO% or more, WBC 4.0 × 10
3
/
µ
L or more)
EO%
75.0% or less, or within ±4.5 EO% (WBC 4.0 × 10
3
/
µ
L or more)
BASO%
120.0% or less, or within ±3.0 BASO%(WBC 4.0 × 10
3
/
µ
L or more)
IG%
75.0% or less, or within ±4.5 IG%
(IG% 2.0% or more, WBC 4.0 × 10
3
/
µ
L or more)
NRBC%
75.0% or less, or within ±4.5NRBC% (WBC 4.0 × 10
3
/
µ
L or more)
NEUT#
24.0% or less (1.20 × 10
3
/
µ
L or more)
LYMPH# 24.0% or less ( 0.60 × 10
3
/
µ
L or more)
MONO#
60.0% or less ( 0.20 × 10
3
/
µ
L or more)
EO#
75.0% or less, or within ±0.36 × 10
3
/
µ
L
BASO#
120.0% or less, or within ±0.18 × 10
3
/
µ
L
IG#
75.0% or less, or within ±0.36 × 10
3
/
µ
L (IG# 0.1 × 10
3
/
µ
L or more)
NRBC#
75.0% or less, or within ±0.36 × 10
3
/
µ
L
Precision
(reproducibility)
Body fluid analysis
mode
WBC-BF 30% or less (CSF: 0.015 - 0.030 × 10
3
/
µ
L, other than CSF: 0.030 - 0.050
× 10
3
/
µ
L)
RBC-BF
40% or less, or Max-Min
≤
0.007 × 10
6
/
µ
L (0.003 - 0.050 × 10
6
/
µ
L)
Analysis parameters
See “Chapter 1: 1.3 Measurement parameters”.
Accuracy (cell counts)
Manual mode,
sampler mode
WBC
Within ±3% or within ±0.20 × 10
3
/
µ
L
WBC-D
Within ±3% or within ±0.20 × 10
/
L
RBC
Within ±2% or within ±0.03 × 10
6
/
µ
L
PLT
Within ±5% or within ±10 × 10
3
/
µ
L
Accuracy (cell counts)
Capillary mode
WBC
Within ±10%
WBC-D
Within ±10%
RBC
Within ±8%
PLT
Within ±8%
Note:
Typical performance characteristics for
precision
(reproducibility) of body fluid analysis mode were
determined during instrument evaluation on a typical instrument, using 14 body fluid samples, counted
10 times consecutively in the body fluid mode and based on the precision profile.
WBC-BF
Reproducibility C.V.25% or less (0.010 – 0.015 × 10
3
/
µ
L)
Reproducibility C.V.20% or less (0.016 – 0.025 × 10
3
/
µ
L)
Reproducibility C.V.15% or less (0.026 – 0.050 × 10
3
/
µ
L)
Reproducibility C.V.10% or less (0.050 × 10
3
/
µ
L or more)
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...