CHAPTER 7
Quality control
Sysmex XE-5000 Instructions for Use
7-13
R
e
vi
se
d A
p
ri
l 20
07
7.7
Change lot
When the change lot is executed, the QC data of the current
lot are deleted, and that of the new lot are moved to the
current lot file.
Click “Change Lot” to display the save confirmation dialog
box for the current lot data.
If saving is necessary, click “Yes (Y)” in the save confirmation
dialog box to store the data in the external storage device.
7.8
Lot No.
Assay values from new lots can be imported by reading the
files from CD-ROM (Read File Function) or by manual entry
via the keyboard.
Information
When the change lot is executed, the QC data in the
current lot file are lost.
When it is necessary to store, save the data to an
external storage device (a USB memory stick or CD-
ROM etc.), then execute the change lot.
Note:
XE-5000 manages the control lots by using two data
files; the current lot and the new lot. The current lot
contains the daily routine Quality Control data, and the
new lot is prepared when the QC material lot is being
switched to the new lot of the control material.
When the control material is to be switched to the new
lot, register in the new lot the lot information by using
the “Lot No.” program (refer to the Section 7.8). Then,
the new control material shall be temporarily run using
both the current lot and the new lot. After examining a
few plots or days, if there is no problem running the
new control material, execute the “Change Lot”
function to move the lot information of the new lot to
the current lot.
Note:
The files which have been set or changed in the Lot
No. dialog box are recorded to the New Lot.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...