
CHAPTER 2
Safety Information
Sysmex XE-5000 Instructions for Use
2-7
R
e
vi
se
d A
p
ri
l 20
07
2.11 Marking on the instrument
Main Unit - Front View
(1)
(2)
(3)
(3)
(1)
(8)
(9)
(6)
(2)
(5)
(7)
(4)
*1
(Inside)
WARNING
Be sure to set the stop bar when opening
to avoid injury.
WARNING
Never touch the detector when the power
of the Main Unit is turned ON. Otherwise,
electrical shock may result.
CAUTION
1. Dye solution replacement method
• For SNR, always replace dye solution and lyse reagent simultaneously.
• For RED, always replace dye solution and diluent simultaneously.
• For FFS, the message “Replace FFS” is displayed after the analysis of
2000 cycles. When doing so, always press Reagent or Confirm after
replacing with new FFS.
2. When the message “Execute Rinse Flowcell” is displayed, always perform
Maintenance – 4. Rinse Flowcell.
3. To clean the IMI detector aperture, first run Maintenance – 1. Drain IMI,
then turn off the main unit power.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...