1. Pressure and Vacuum Errors
Note:
5. Emergency stop
If an error occurs, the analysis operations will
immediately be discontinued, and all the sequences
will stop.
Then the message prompting you to turn OFF the
power will appear. To restore the system, turn the
power OFF, wait for at least 10 seconds, and then
turn it ON again.
• Sample data with an error will be displayed in “----.”
Error messages:
0.25 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. 0.25 MPa pressure adjustment error
2. Inadequate pressure in the Pneumatic Unit
3. Pressure leakage from tube or nipple
Corrective Action
1. Adjust pressure of 0.25 MPa. (Refer to Chapter 9: 9.21
Adjusting pressure and vacuum.)
2. Check the Pneumatic Unit power switch and the power cord
connection.
3. Check the tube connection and nipple looseness.
Error messages:
0.16 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. 0.16 MPa pressure adjustment error
2. Defective 0.16 MPa regulator
Corrective Action
1. Adjust pressure of 0.16 MPa. (Refer to Chapter 9: 9.21
Adjusting pressure and vacuum.)
2. If adjustment of 0.16 MPa pressure fails, the regulator is
probably defective. Contact your Sysmex technical
representative.
Error messages:
0.07 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. 0.07 MPa pressure adjustment error
2. Defective 0.07 MPa regulator
Corrective Action
1. Adjust pressure of 0.07 MPa. (Refer to Chapter 9: 9.21
Adjusting pressure and vacuum.)
2. If adjustment of 0.07 MPa pressure fails, the regulator is
probably defective. Contact your Sysmex technical
representative.
Error messages:
0.03 MPa ERROR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. 0.03 MPa pressure adjustment error
2. Defective 0.03 MPa regulator
Corrective Action
1. Adjust pressure of 0.03 MPa. (Refer to Chapter 9: 9.21
Adjusting pressure and vacuum.)
2. If adjustment of 0.03 MPa pressure fails, the regulator is
probably defective. Contact your Sysmex technical
representative.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...