CHAPTER 9
Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use
9-49
R
e
vi
se
d A
p
ri
l 20
07
E
U
4. Check that the pressure displayed on the Sensor 1 screen
is the predetermined pressure.
5. If the displayed pressure does not match the
predetermined pressure, adjust it by repeating the
operations 3 and 4.
6. When the adjustment is completed, press the adjustment
knob for the 0.07 MPa regulator to lock.
Adjusting the pressure to 0.03 MPa
This pressure is supplied to RBC/PLT Sheath detector.
1. Display the Sensor 1 screen on the LCD screen.
See “Chapter 9: 9.21 Displaying the pressure or vacuum”.
2. Pull out the knob for the 0.03 MPa regulator to release the
lock.
3. Adjust the pressure while watching the pressure displayed
on the Sensor 1 screen.
To increase the pressure, turn the knob clockwise.
Normal range: 0.03 ± 0.001 MPa
4. Check that the pressure displayed on the Sensor 1 screen
is the predetermined pressure.
5. If the displayed pressure does not match the
predetermined pressure, adjust it by repeating the
operations 3 and 4.
6. When the adjustment is completed, press the adjustment
knob for the 0.03 MPa regulator to lock.
Information
Always adjust pressure so as to increase to the
predetermined level. If the pressure is too high, lower
it below that level once; then raise it to the set level.
Information
Always adjust pressure so as to increase to the
predetermined level. If the pressure is too high, lower
it below that level once; then raise it to the set level.
Adjustment
knob
Low
High
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...