11.7 Check before installation
The XE-5000 and associated equipment is installed by your
Sysmex technical representative. In case relocation becomes
necessary after installation, contact your Sysmex technical
representative.
Problems resulting from the relocation of the XE-5000 by
anyone other than a Sysmex technical representative are not
covered by the Warranty even if it is in the warranty period.
11.8 Grounding
The instrument power supply cord uses a 3-prong plug. When
the power supply socket is provided with grounding, simply
plug it to the socket. If the power supply socket is not provided
with grounding, use an adapter plug and ground the ground
wire separately.
11.9 Installation environment
• Use sample at an ambient temperature of 15 – 30ºC (20 –
30ºC for HPC analysis mode).
• Relative humidity should be within the range of 30% - 85%.
• If ambient temperature and relative humidity are not within
the suggested range, air-condition the environment.
• Avoid a place that can be extremely high or low in
temperature.
• Avoid a place that is exposed to direct sunlight.
• Select a well-ventilated place.
• Avoid a place close to a centrifugal machine, wireless
communication devices or communication facility where
high frequency waves are generated or radio interference
can occur.
Warning!
• Always ground this instrument.
Inadequate grounding creates the danger of
electrical shock.
• Take care to stay within the capacity of the socket.
This is to prevent the risk of a fire.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...