
CHAPTER 9
Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use
9-39
R
e
vi
se
d A
p
ri
l 20
07
E
U
Removing the piercer
1. Turn OFF the Main Unit power and open the front cover.
2. Loosen the fixing screws and remove the CP cover.
3. Remove the fixing screws of the piercer cover and remove
the cover.
4. Remove the tubes and rubber joint which are connected
to the rinse cup and piercer.
5. Install the fixing board to the piercer using three screws.
Information
Another tube is installed inside one tube. The
replacement part of this inner tube is attached to the
new piercer set. Remove the inner tube together with
the outer tube.
Warning!
Be sure to install the piercer fixing board.
The piercer needle may pop out during operation and
cause injuries.
Loosen
Fixing screw
CP cover
Loosen
Fixing screw
Piercer cover
Rubber joint
Rinse cup
Piercer
Tube
Piercer fixing board
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...