
CHAPTER 10
Troubleshooting
Sysmex XE-5000 Instructions for Use
10-13
R
e
vi
se
d A
p
ri
l 20
07
Error messages:
FCM RU Temp High
FCM RU Temp Low
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of reaction chamber is outside the regulation
range.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range.
If this error is still displayed 30 minutes after the power is
turned ON, there is probably something defective in the
system. Contact your Sysmex technical representative.
Error messages:
FCM Detector Temp High
FCM Detector Temp Low
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. Temperature of optical detector block is outside the
regulation range.
Corrective Action
1. Wait until the temperature has stabilized inside the
regulation range.
If this error is still displayed 30 minutes after the power is
turned ON, there is probably something defective in the
system. Contact your Sysmex technical representative.
Error messages:
IMI TD Therm Sens ERR
Status:
Not Ready
Probable Cause
1. One of the thermal sensors for IMI detector has broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Error messages:
IMI TD Therm Sens ERR HPC
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. One of the thermal sensors for IMI reaction chamber has
broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative. In this case,
it is possible to perform analysis in normal mode even if the
error persists.
Error messages:
33°C RH Therm Sens ERR
Status:
Not Ready
Probable Cause
1. One of the thermal sensors for 33ºC reagent heater has
broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Error messages:
40°C RH Therm Sens ERR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. One of the thermal sensors for 40ºC reagent heater has
broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Error messages:
FCM RU Therm Sens ERR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. One of the thermal sensors for reaction chamber has
broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...