CHAPTER 9
Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use
9-13
R
e
vi
se
d A
p
ri
l 20
07
7. Loosen the detector chamber fixing screw, and remove
the chamber.
8. Apply CELLCLEAN on the brush provided, and dab by
lightly prodding the brush against the aperture.
9. Install the detector chamber to its original position, and
tighten the fixing screws.
10. Close the detector cover and Main Unit front cover.
11. Turn ON the power of the Main Unit.
12. Background check starts automatically. Make sure that all
background values are within acceptable.
Information
When removing the detector chamber, take care not to
apply excessive force to the tube that is connected to
the detector chamber.
Warning!
CELLCLEAN is a strong alkaline detergent. Pay
careful attention not to let it adhere to your body or
clothes. If CELLCLEAN adheres to your body or
clothes, wash it away immediately with plenty of water.
Otherwise, it can damage your skin or clothes.
Caution!
Be sure to use CELLCLEAN only.
Note:
After using the brush, wash it in water thoroughly to
remove CELLCLEAN before storing it.
Caution!
When installing the detector chamber, make sure that
the O-ring is placed in the specified position.
Tighten the fixing screws alternately and evenly.
Otherwise, analysis may not be performed correctly.
Detector chamber fixing screw
Loosen
Aperture
Brush
O-ring
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...