CHAPTER 10
Troubleshooting
10-36
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
Motor
1. WB Aspiration Motor
This program performs test operation of the WB aspiration
motor and sets the motor in the correct initial position.
Execute this test when a “WB Asp Motor Error” occurs and
analysis cannot be performed. If the test is normally
completed, analysis will be possible after the system enters
the READY status.
1. From the Main Menu screen on the Main Unit, select
“Test” on the function menu.
The Test Menu screen will appear.
2. Select “Motor” on the function menu of the Test Menu screen.
The Motor Test screen will appear.
3. Select “1. WB Aspiration Motor” by moving the cursor with
or
key on the Motor Test screen.
4. Select “Execute” on the function menu of the Motor Test
screen. The WB aspiration motor will start running, and
the Testing screen will appear.
5. When aspiration is completed, the motor will return to the
initial position.
6. To cancel the motor test, select “Cancel” on the function
menu, or press the RETURN key.
2. RBC Sheath Injector
This program performs test operation of the RBC sheath
injector and sets the motor in the correct initial position.
Execute this test when a “RBC Sheath Motor Error” occurs
and analysis cannot be performed. If the test is normally
completed, analysis will be possible after the system enters
the READY status.
1. From the Main Menu screen on the Main Unit, select
“Test” on the function menu.
The Test Menu screen will appear.
2. Select “Motor” on the function menu of the Test Menu
screen.
The Motor Test screen will appear.
3. Select “2. RBC Sheath Injector” by moving the cursor
with
or
key on the Motor Test screen.
4. Select “Execute” on the function menu of the Motor Test
screen. The RBC sheath injector will start running, and
the Testing screen will appear.
5. When the analysis operations are completed, the motor
will return to the initial position.
6. To cancel the motor test, select “Cancel” on the function
menu, or press the RETURN key.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...