CHAPTER 6
Operation
Sysmex XE-5000 Instructions for Use
6-7
Revised July
2
007 EU
●
Samples Analyzed per Reagent Container
• The number of cycles per container refers to the number of
cycles that can be run in CBC+DIFF+RET+NRBC analysis
mode.
●
Replacing the Reagent
After replacement of the reagent, perform automatic cleaning,
check that the background count is low, then start the sample
analysis.
Reagent
Abbreviation No. of cycles/
container
Capacity
per
container
CELLPACK
EPK
Approx. 660
cycles
20.0 L
CELLSHEATH
ESE
Approx. 7300
cycles
20.0 L
STROMATOLYSER-
FB
FBA
Approx. 2750
cycles
5 L
STROMATOLYSER-
4DL
FFD
Approx. 2750
cycles
5 L
STROMATOLYSER-
4DS
FFS
Approx. 2000
cycles
42 mL
STROMATOLYSER-
NR(L)
SNR
Approx. 2000
cycles
3.6 L
STROMATOLYSER-
NR(S)
Approx. 2000
cycles
43 mL
SULFOLYSER
SLS
Approx. 10000
cycles
5 L
STROMATOLYSER-
IM
SIM
Approx. 3200
cycles
10 L
RET SEARCH (II)
(Diluent)
RED
Approx. 550
cycles
1 L
RET SEARCH (II)
(Dye solution)
Approx. 550
cycles
12 mL
Caution!
• Use reagent that has been left at room temperature
(15–30°C) for at least 24 hours.
• In handling a reagent that may have frozen, follow
the precautions given on the package.
Otherwise analysis may not be performed correctly.
• When replacing the reagent container, make sure
that no dust adheres to the cubitainer spout kit.
Otherwise analysis may not be performed correctly.
• After opening a reagent container, make sure that no
dust, dirt or bacteria comes into the container.
Otherwise analysis may not be performed correctly.
Содержание XE-5000
Страница 2: ......
Страница 42: ...CHAPTER 3 Design and Function 3 18 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 256: ...CHAPTER 11 Technical Information 11 38 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Страница 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...