![Svantek SV 258PRO Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/svantek/sv-258pro/sv-258pro_user-manual_1409264080.webp)
SV 258 PRO (AG) User Manual
80
In case of the
User
standard and 1/3-octave spectrum
(
Octave 1/3 Vel.
or
Octave 1/3 Acc.
) if
Curve
is
selected as a
Source
, two additional positions
Condition 1
and
Condition 2
appear.
These positions allow you to set additional conditions
that must be met to trigger an event which is based on
comparing the RMS values from two selected profiles.
After enabling the additional condition, you should
select two results for different profiles that will be
compared: RMS with RMS or rolling RMS (
RRMS
) with
rolling RMS.
The condition will be met when the left result is greater
than the right result.
The selected threshold condition will be presented in
the TRIGGER CONDITIONS pop-up box.
In this box you can delete this condition clicking the
icon or confirm the selection clicking
OK
.
After confirmation the pop-up box closes, and the
selection will be presented in the line of the
TRIGGER
button.
Defining actions
To create new action, click the
field and in the
ADD EVENT ACTION
pop-up box, click the action
you wish to add and to configure:
Alarm lamp
,
SMS alarm
or
E-mail alarm
.