Prism6 & Crystal Reader software User Manual
15
•
96 well plate tray holder
– H24003, to be used for Prism6 instrument service maintenance
operation only. Only to be used by Stilla Technologies Service Specialist.
Up to 3 chips can be fitted into one tray holder for scanning at a time.
The direction for insertion of the tray holders onto the moveable tray is indicated on the individual
tray holder labels. Stilla Technologies recommends cleaning all tray holders occasionally with
70% ethanol solution, rinse with water using a lint-free wipe.
Position one of the chips (Sapphire chips or Opal chips) in the individual slots of the
corresponding tray holder (Sapphire tray holder or Opal tray holder). Check for correct
positioning. The respective chips should fit perfectly into the individual slots.
Instructions to position the chips in the tray holder.
Example provided for Sapphire chips and Sapphire tray holder only. Similar instructions
apply for Opal chip placement in Opal tray holder. Differences are indicated in the provided
text description with the respective image instructions.
Place the tray holder on a clean surface.
Release the chip clamp by rotating it away
from the slot.