Prism6 & Crystal Reader software User Manual
14
a)
Press the
“OPEN TRAY” button on the Crystal Reader software
homepage.
b)
Alternatively, p
ress the “Eject” button located at the front of the Prism6
instrument (Fig 1).
Wait until the moveable tray is fully ejected before placing the chips for scanning.
Figure 3:
Fully ejected position of the movable tray. In this position, the moveable tray is ready
to be loaded with the tray holder for chip scanning.
CAUTION! Do not manually open the Prism6 instrument flap. Do not manually engage with
the movable tray. Do not manually pull out the moveable tray any further.
c. Instructions for chip positioning
Note:
Prior to chip positioning for scanning read the respective Sapphire chip and Opal chip
Instructions for Use for detailed instructions on how to clean the bottom surface of each chip
and for automatic chip ID barcode reading using a handheld USB barcode reader.
Figure 4 -: Sapphire tray holder (left) & Opal tray holder (right)
Three kinds of tray holders are provided with each Prism6 instrument:
•
Opal tray holder
– H24001, to be used with Opal chips only
•
Sapphire tray holder
– H24002, to be used with Sapphire chips only