Prism6 & Crystal Reader software User Manual
17
Correct positioning of a chip in the tray
holder.
Incorrect positioning of a chip in the tray
holder. The chip is lying at an angle in the
tray holder and is not well attached to the
top-right corner of the tray holder.
Repeat the described chip positioning operation for all the tray holder slots to be used.
Note: Make sure to use the specific Sapphire tray holder or Opal tray holder with the respective
chip types only. Do not operate Sapphire chips on Opal tray holder. Do not operate Opal chips
on Sapphire tray holder. Do not mix chip types on one tray holder. The tray holder is clearly
identifiable by the tray holder label
.
Instructions for loading the tray holder in the movable tray of the Prism6 instrument:
Example provided for Sapphire chips and Sapphire tray holder only. Identical instructions
apply for Opal tray holder placement.