STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10A-M Gas Detector
TD18/032
Issue:
10
Change Note:
1869
Page 62
9.1.26.
Set Flammable Gas Correction Factor
NOTE: This option is for use by SSCL only and is password protected.
It is used to set the Flammable gas reading, corrected for Propane, when
using a propane sensor.
Use the keys Select the desired flammable gas and select ENTER
The current correction factor is then displayed above the gas name.
SetFlam .Correction
Edit
Chloromethane
4.97
Use the keys to change the digit. Select the next digit using the
keys.
•
Press ENTER to accept the new value will be
displayed.
Note: Pressing MENU instead of ENTER leaves the settings
unchanged.
•
Press the MENU button to exit the sensor type option menu.
On exit the display will show if no changes have
been made or if changes have been made.
There is also a User Defined option to cater for any flammable gases not on
the list which have a usable propane response. The correction factor (and
LEL) may be set by the user.
SetFlam.Correction
Edit
User Defined
1.00
No change
Data updated
Updated