STATUS SCIENTIFIC CONTROLS
Installation, Commissioning & Routine Gas Testing
FGD10A-M Gas Detector
TD18/032
Issue:
10
Change Note:
1869
Page 28
9.1.2. Sensor Zero
•
From the menu system select menu option:
Sensor Zero
and
press ENTER.
Calibration
Zero Sensor
•
Ensure the sensor is in a zero-gas environment.
Note:
1. Carbon Dioxide sensors cannot be zeroed in air due to the
background levels of CO
2
present. These sensors are best
zeroed whilst being exposed to 100% nitrogen.
2. Where a purging gas has to be applied, use a flow rate of
between 500 and 1000cc/min. Allow sufficient time for the
sensor to respond.
•
Press ENTER to zero the sensor, will be displayed
to confirm the sensor zero has been performed.
Note:
Pressing MENU rather than ENTER exits the zero feature
without performing the calibration.
•
Press MENU to close the menu system.