77
:SOUR:SAFE:STEP 1:IR:LIM:HIGH 1000000.
--Example:
:SOUR:SAFE:STEP 1:IR:LIM:HIGH?
If the current UPPER resistance for step 1 is 1000000,
1000000 is returned.
:SOUR:SAFE:STEP:IR:TIME:RAMP
Sets the RISE time for IR test.
Also inquires about the RISE time.
--Syntax:
Command message:
SOUR:SAFE:STEP <sn>:IR:TIME:RAMP <time>
Query message:
SOUR:SAFE:STEP <sn>:IR:TIME:RAMP?
--Program data<sn>:
Data format: Integer
Set value: 1~49
Resolution: 1
Unit:
--Program data<time>:
Data format: float
Set value: 0~999.9 (0 is OFF)
Resolution: 0.1
Unit: s
--Example:
To set the RISE time for step 1 to 1,
:SOUR:SAFE:STEP 1:IR:TIME:RAMP 1.
--Example:
:SOUR:SAFE:STEP 1:IR:TIME:RAMP?
If the current RISE time for step 1 is 1,
1 is returned.
:SOUR:SAFE:STEP:IR:TIME:FALL
Sets the FALL time for IR test.
Also inquires about the FALL time.
--Syntax:
Command message:
SOUR:SAFE:STEP <sn>:IR:TIME:FALL <time>
Query message: