17
3.2 Instruction of rear panel
WARNING:
FOR CONTINUED PROTECTION AGAINST FIRE
HAZARD,REPLACE ONLY WITH THE SAME TYPE AND RATING OF FUSE
AS SPECIFIED FOR THE LINE VOLTAGE BEING UTILIZED.
CAUTION:
NO OPERATOR SERVICEABLE PARTS INSIDE
REFER SERVICING TO QUALITIFIED PERSONNEL.
90V-110V~
108V-132V~
198V-242V~
216V-250V~
5AT
250V
2.5AT
250V
HIGH VOLTAGE
MAX:5kV AC 30mA/6kV DC 10mA
HV OUTPUT
RTN/
LOW
100V/120V/220V/240V~
50/60Hz 500W MAX
GPIB
CONT.
CHECK
OPTION
1Ω CAL
VOLTAGE SELECTOR
POLLUTION DEGREE 2
INSTALLATION CATEGORY Ⅱ
贴标签
!
!
OUTPUT
FAIL
START
RESET
COM
INTER
LOCK
UNDER
TEST
PASS
1
5
9
UNDER
RESET
TEST
START
PASS
FAIL
6
SCANNER
RS-232C
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Figure 3-2
1.
Fan
Power amplifier circuit radiator.
2.
Line voltage range
Switch of input voltage range.
3.
PLC controller interface
Used to connect programming controller, where:
INTERLOCK: input the connecting locked signal, if off, starting output is not
allowed.
TEST:
Output synchronized-signal control when high voltage output is
started.
START: Input the starting signal for outputting high voltage, corresponding
to the START button on the front panel.
RESET: Input the reset signal for stopping output of high voltage,
corresponding to the STOP button on the front panel.
PASS:
Indicate the test result, corresponding to PASS indication on the
front panel.
FAIL:
Indicate the test result, corresponding to FAIL indication on the
front panel.
4.
OUTPUT
TEST output control, output 24V level or potential-free relay switch signal; it is
suggested to use the switch signal to assure signal separation.
5.
HANDLER interface
Use 9 core model D jack to output.
6.
Ground break-over test interface
When open ground break-over test function, the terminal has to be connected to