91
IR test
Output voltage
0.050V – 1.500V
Voltage resolution
1V
Voltage test accuracy
±(1.0% 2V)
Max. output current
10mA
5mA
Max. output power
10 VA (1000V/1mA)
5VA
(1000V/1mA)
Output short current
≥20mA
≥10mA
Load regulation
≤1
%
(nominal power)
Ripple (1kV)
≤3% (1kV, non-load)
Discharge function
Auto discharge after test ends
Resistance
measurement range
0.01MΩ– 50GΩ, (current range: 1nA – 10mA)
Resistance display
range
(1000V)
10 mA
0.1 MΩ-10 MΩ
3mA
0.3 MΩ-30 MΩ
300µA
3 MΩ-300 MΩ
30µA
30 MΩ-3GΩ
3µA
300 MΩ-30GΩ
300nA
>30GΩ
Resistance
measurement accuracy
≥500V
1MΩ–1GΩ
±(5% r5 digits)
1GΩ–10GΩ
±(10% r5 digits)
10GΩ–50GΩ
±(15% r5 digits)
<500V
0.1MΩ–1GΩ
±(10% r5 digits)
Current measurement
accuracy
±(1.5% r5 digits) (after OFFSET
correction)
Arc detection
Measurement
range
AC
1mA – 3mA
1mA – 3mA
DC
1mA – 10mA
1mA – 5mA
------
Comparator
Judgement mode
Window comparison mode:
I
lower
ON: at I
lower
< Ix < I
upper
,
PASS; (condition:
at I
x
≤ I
lower
or I
x
≥ I
upper
, FAIL
I
lower
< I
upper
)
I
lower
OFF: at I
x
< I
upper
, PASS; at I
x
≥ I
upper
, FAIL
The same judgment mode for insulation resistance judgment.
Current
upper limit
I
upper
AC
0.001mA – 30mA
0.001mA – 20mA
DC
0.1µA – 10mA
0.1µA – 5mA
------
Current lower
limit I
lower
(LOWER OFF)
AC
0.001mA – 30mA
0.001mA – mA
------
DC
0.1µA – 10mA
0.1µA – 5mA
------
Resistance upper limit
OFF - 0.01MΩ - 50GΩ
------
Resistance lower limit
0.01MΩ– 50GΩ
------
Judging output
PASS/FAIL LCD or LED respectively display, sound
alarm
------