7
2. On the rear panel, remove the fuse holder, by pushing it inward and
unscrewing it counterclockwise using a screwdriver.
3. In accordance with the fuse rating specified below, check the fuse type and
replace the fuse.
4. Following the above steps in the reverse order, reinstall the fuse holder.
Voltage range
Frequency
Fuse type
Power
100V, 120V
47-63Hz
5A(ST9201/S)
500VA
3A(ST9201B/C)
350VA
220V, 240V
3A(ST9201/S)
500VA
2A(ST9201B/C)
350VA
1.5 Connecting the AC Power Cord
The power cord that is provided varies depending on the destination for the
product at the factory-shipment.
Do not use the AC power cord provided with the product as an AC
power cord for other instruments.
Connection procedure:
1. Confirm that the supply voltage is within the line voltage range of the tester.
2. Confirm that the line power are correctly selcected.
3. Confirm that the POWER switch on the tester is off.
4. Connect the AC power cord to the AC LINE connector on the rear panel.
5. Use the provided power code or power code that is selected by qualified
personnel.
6. Plug in the AC power cord.
1.6 Grounding
!
WARNING
Be sure to connect the tester to an electrical ground (safety
ground). If the output to a conveyer or peripheral device that is connected to
an earth ground or a nearby commercial power line is short-circuited without
grounding, the tester chassis is charged to an excessively high voltage,
resulting in extreme danger.
This tester is designed as a Class I equipment (equipment protected against
electric shock with protective grounding in addition to basic insulation).
Therefore, electric shock may occur without proper grounding.
To ensure safety, be sure to ground the tester.