◇
88
Chapter 6 ST9201 series
specification
Model
ST9201
ST9201S
ST9201B
ST9201C
Withstanding voltage test
Output
voltage
A
C
Range
0.050kV—5.000kV
Waveform
Sinusoidal wave
Distortion
< 3%
Frequency
50, 60Hz selectable
Frequency
accuracy
±2%
Max. output
power
150VA (5.000kV 30mA)
100VA (5.000kV 20mA)
D
C
Range
0.050 kV—6.00kV
------
Signal
source
frequency
600Hz
------
Max. output
power
50VA
(5.000kV 10mA)
25VA
(5.000kV
5mA)
------
Voltage
regulation
±(1.0% +10V) (nominal power)
Voltage
resolution
1V
Voltage display
accuracy
±(1.0%5V) (non-load)
Voltage
generating
method
DDS signal AB power amplifier
Current
test
range
A
C
Range
0.001mA – 30.00 mA
0.001mA – 20.0mA
Short-circuit
current
>60 mA (output
voltage>500V)
>40 mA (output
voltage>500V)
Resolution
0.001 mA
0.001 mA
Accuracy
±(1% r5 digits)
±(1% 5 digits)
Effective
current
OFF-0.001 mA-30 mA
OFF-0.001 mA-20 mA
D
C
Range
0.1µA – 10.00mA
0.1µA – 5.00mA ------
Accuracy
±(1.0% r5 digits)
±(1.0% reading
+5 digits)
------
Discharge
function
Auto discharge after test ends (DCW)
------