:ALLSET3 <func>,<scan>,<open>,<shrt>,<stan>
5 items, see parameters list.
--Response message:
‘1’ if the command was valid.
Parameters list:
Items
Type
Function
Detail
func
enum
Setup functions
0--None; 1--AC; 2--DC; 3--IR; 4--OS
scan
int
Setup channel
Bit mask for channel selection, see Note1
volt
float
Setup voltage
AC:50~5000,DC:50~6000,IR:50~1000
upper
float
Setup upper limit
Unit: µA
lower
float
Setup lower limit
Unit: µA
time
float
Setup test time
0.3~999.9s
arc
float
Setup arc level
Unit: µA
real
float
Setup real current
Unit: µA
freq
enum
Setup frequency
0—50Hz ,1—60Hz
rise
float
Setup rise time
0~999.9s
fall
float
Setup fall time
0~999.9s
wait
float
Setup wait time
0~999.9s
sagc
enum
Setup soft AGC
0—OFF,1--ON
check
enum
Setup GND check
switch
0—OFF,1--ON
range
enum
Setup range
0—Auto,1~6 hold range
open
enum
Setup open level
0—10%,1—20%......10—100%
shrt
enum
Setup short level
0—OFF,1—100%,2—200%......5—500%
stan
float
Setup standard value
Unit: nF
offset
float
Setup offset value
Unit: µA
Note 1:
The value of scan is a 16-bit mask with the individual bits defined as follows:
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0
CH1
_L
CH2
_L
CH3
_L
CH4
_L
CH5
_L
CH6
_L
CH7
_L
CH8
_L
CH1
_H
CH2
_H
CH3
_H
CH4
_H
CH5
_H
CH6
_H
CH7
_H
CH8
_H
CHn_L
0
1
0
1
CHn_H
0
0
1
1
Settings
Open
High
Low
Error