27
There are mainly 4 function keys which are SETUP, TEST, SYSTEM, MEMORY.
4.2.1
SETUP
STEP:
01/01
AC
SCAN:
1X2X3X4X5X6X7X8X
VOLT:
0.050
kV
RISE:
0.5
S
LOCK
UPPER:
1.000
mA
FALL:
0.5
S
OFST
LOWR:
OFF
REAL:
OFF
RMT
TIME:
0.5
S
FREQ:
50
Hz
ERR
ARC:
OFF
INS
DEL
NEW
F1
F2
F3
F4
F5
Figure 4.2.1 AC setup interface
Instruction:
STEP: 01/01
Test procedure: current setting No. / total items.
Key
Function Instruction
F1
INS
Within the current test programme (PROG), it adds a
new test item.
F2
DEL
Within the current test programme (PROG), delete
the current test item.
F3
NEW
Create a blank test programme (PROG)(including a
new test item), which is used to write a new test
programme.
ENTRY
Switch the current selected parameters.
START
It can be used to switch to the TEST interface in
SETUP interface and start high voltage test.
AC Current operating mode of test procedure is AC withstanding voltage.
Cursor in this position can switch to other working items by F1~F3, for instance DC,
IR, OS.
SCAN: 1X2X3X4X5X6X7X8X Built-in multichannel output and test terminal.
SCAN channel: built-in multichannel circuits
SBOX: 1X2X3X4X5X6X7X8X External multichannel output and test
terminal. (custom)
SBOX channel: external ST9121 multichannel output port
The output interface of the multichannel controller can be set to connect with the
high voltage output terminal, test terminal and hanging open.
The connection of 8 output channels and test output terminals:
The channel values are:
Open (x), high voltage output (
■
)and test low terminal (
□
).
Other data
The test condition of current (AC) item (in 4.3.1 AC
withstanding voltage setup).