35
4.3
Test item interface and parameter setup
4.3.1
AC withstanding voltage test parameter setup:
STEP:
01/01
AC
SCAN:
1X2X3X4X5X6X7X8X
VOLT:
0.050
kV
RISE:
0.5
S
LOC
UPPER:
1.000
mA
FALL:
0.5
S
OFST
REAL:
OFF
LOWR:
0.001
mA
RMT
TIME:
0. 5
S
FREQ:
50
Hz
ERR
ARC:
OFF
F1
F2
F3
F4
F5
Figure 4.3.1 AC setup interface
Instruction of AC test parameters:
VOLT:
0.050~5.000kV
Voltage value of AC high voltage
test
UPPER:
0.001~30.00mA
Current high limit value of AC
withstanding voltage
REAL:
OFF~0.001~30.00mA Real current high limit value of AC
withstanding voltage
TIME:
OFF~0.1~999.9S
Test time of AC withstanding
voltage
ARC:
OFF~0.1~15.0mA
Current Max. value of AC arc
RISE:
OFF~0.1~999.9S
Voltage rising time of AC high
voltage test
FALL:
OFF~0.1~999.9S
Voltage falling time of AC high
voltage test
LOWR:
OFF~0.001~30.00mA Low limit current value of AC
FREQ:
50/60
AC working frequency
4.3.2
DC withstanding voltage test parameter setup:
STEP:
01/01
DC
SCAN:
1X2X3X4X5X6X7X8X
VOLT:
0.050
kV
RISE:
0.5
S
LOC
UPPER: 1.000
mA
FALL:
0.5
S
OFST
LOWR:
OFF
WAIT:
OFF
RMT
TIME:
0. 5
S
CHEK:
OFF
ERR
ARC:
OFF
F1
F2
F3
F4
F5
Figure 4.3.2 DC setup interface
Instruction of DC test parameters:
VOLT:
0.050~6.000kV
Voltage value of DC high voltage
test
UPPER:
0.1µA~10.00mA
Current high limit value of DC