4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.7 Stand-alone water supply
båÖäáëÜ
In the
Start
sub-screen, the
Aqua
key is highlighted in orange.
The disinfection system is now changed to stand-alone water
supply.
Resetting disinfection system for operation with public drinking water
and disinfectants
Resetting the disinfection system to public drinking water is similar to
changing to stand-alone water supply. Instead of the
Refilling water
display, the
Refilling disinfectant
display appears on the touchscreen.
1. Fill approx. 0.5 liters of DENTOSEPT P into the disinfectant tank.
2. Sanitize the water lines after switching to the public drinking water
supply (see “Sanitizing [ → 190]”).
Level control
If the
Refill water
or
Refill disinfectant
display appears during operation,
the tank is almost empty (< 300 ml). Pressing the
Return
key hides the
display and allows the operation to continue. Refill the tank as soon as
possible.
Water unit, disinfectant tank < 300 ml
Setup
San
Aqua
Apex Dist
Auto
Setup
San
Aqua
Auto
Apex Dist
Stop
Desinf
IMPORTANT
300 ml are approximately sufficient for:
➢ 5 min. spray operation (motor, turbine, SIROSONIC L) or
➢ 2 min. SPRAYVIT operation
The actual values depend on the instrument types and settings used.