3 Unit description
Sirona Dental Systems GmbH
3.6 Dentist element
Operating Instructions Simulation Unit SIM MODULAR
Endodontics/Purge
With the motor removed: switching the integrated torque control for motor
BL ISO E on/off, see "Switching integrated torque control ON/OFF"
[ → 79]
With the ultrasonic handpiece removed: switching the endo mode on/off
(limiting power), see "Setting the intensity on the EasyPad" [ → 90].
With motor removed and torque control switched on: activating the
calibration process (> 2 s), see "Calibrating the burr drive" [ → 80]
With instruments in place: selecting the purge, autopurge, sanitize, and
stand-alone water supply functions
Display mode/clean
Display mode/clean
With motor removed and torque control switched on: Switching between
torque and speed display, see “Setting the speed” [ → 83] in the section
“Integrated speed control”.
With instruments in place: Deactivates the complete user interface of the
dentist element. The instruments of the dentist and assistant elements
can no longer be activated. The SIVISION monitor and the camera
system are switched off. Pressing it again > 3 s reactivates the simulation
unit. This is used to clean the surface and protect against interference
from an external HF surgical unit so that no unwanted functions can be
accidentally triggered, see “Disinfecting user interfaces”. Electromagnetic
interference can often be reduced, if the external HF surgical unit is
operated with a neutral electrode.
Function key
Starts and stops timer, see "Timer function".
Purge
Clean
Fn