65 64 020 D3650
D3650
.
201.01.02
.
02 04.2016
19
Sirona Dental Systems GmbH
3 Unit description
Operating Instructions Simulation Unit SIM MODULAR
3.1 Technical data
båÖäáëÜ
3
Unit description
3.1
Technical data
Model designation, power ratings
Model designation
SIM MODULAR simulation workstation
Power supply
230 V AC ± 10%
50/60 Hz
Rated current
1.5 A at 230 V AC
Additional max. of 6 A for external devices
Main building fuse
Media
Power ratings
Operating conditions
Type B automatic circuit
breaker
Slow-blow 100–115 V AC:
20 A
Slow-blow 220–240 V AC:
16 A
Overvoltage category
II according to IEC 60664-1
Protection class
I
Water quality
Drinking water
Water hardness
1.4 mmol/l (=8°dH) -2.1 mmol/l (=12°dH)
Air quality
Dry, oil-free, dirt-free, not contaminated.
Average power consumption
(for dimensioning an air
conditioning system):
100 W
Power consumption in
switched off state
0 W (main switch available)
On-site water pressure
2.5-6 bar ( ≥ 3 l/min)
On-site air pressure
5.5-7.5 bar ( ≥ 50 l/min)
On-site negative pressure of
suction air
P ≤ 0.18 bar ( ≥ 500 l/min)
Waste water:
max. 3 l/min
Intended use
For indoor use
Ambient temperature
10°C-40°C (50°F-104°F)
Relative humidity
30%-85%
Air pressure
500hPa-1060hPa
Installation site
≤ 3000 m above sea level
The simulation unit is not suitable for
operation in explosion hazard areas.
Pollution degree
2 acc. to IEC 60664-1