5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.10 Sanitizing
Operating Instructions Simulation Unit SIM MODULAR
Canceling the sanitizing process
Sanitation can be canceled during some sanitation phases, e.g. if the
simulation unit must be switched back on before the 24 hours have
expired. The
Stop
key shows whether cancellation is possible.
✔
The sanitizing process has started.
➢ Touch the
Stop
key on the touchscreen.
The sanitation process is canceled. If the water lines have not yet
been treated with DENTOSEPT P disinfectant (prior to sanitation
phase 3), the simulation unit automatically proceeds directly to
sanitation phase 6. If the sanitation process is canceled during
the reaction time, the water lines are initially rinsed with water
(sanitation phase 5).
Cleaning the suction hoses
After sanitization, all suction hoses should also be chemically cleaned:
● If the simulation unit is not equipped with the chemical suction hose
cleaning option, see “Suction hose cleaning without “suction hose
● If the simulation unit is equipped with the chemical suction hose
cleaning option, see “Suction hose cleaning with “suction hose
Stop