4 Operation
Sirona Dental Systems GmbH
4.5 Dentist element
Operating Instructions Simulation Unit SIM MODULAR
4.5.6
General instrument functions
EasyPad
Settings for the coolant, instrument light, and foot switch can be made for
the instrument removed from the holder.
EasyTouch
Settings for the coolant, instrument light, and foot switch can be made in
the sub-screen of the instrument removed from the holder.
The sub-screens vary according to the instrument removed. Functions
not available for the respective instrument are not displayed in the sub-
screen.
4.5.6.1
Instrument functions on the EasyPad
4.5.6.1.1
Switching spray ON/OFF
The spray is switched on when the foot switch is activated in combination
with the instrument. On the EasyPad, the spray is switched on/off via the
foot switch.
Cooling of treatment areas
Flow rate reduced
Switch spray on/off with the pneumatic foot switch
With a pneumatic foot switch
➢ Toggle the switch on the pneumatic foot switch.
Left position: Spray off
Right position: Spray on
The spray is switched on/off when the instrument is activated.
Activation of the spray is not displayed on the EasyPad when a pneumatic
foot switch is used. The status display
Spray
does not light up.
CAUTION
Instrument can be operated without coolant.
Tooth substance can be damaged by frictional heat.
➢ Always make sure that the treatment area does not overheat
whenever you switch the coolant off.
CAUTION
If air and water are taken from an external connection, the flow rate in
the instruments can drop.
Tooth substance can be damaged by frictional heat.
➢ Avoid the removal of water from additional devices during treatment
with treatment center instruments.