65 64 020 D3650
D3650
.
201.01.02
.
02 04.2016
79
Sirona Dental Systems GmbH
4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
4.5.10.1
Integrated torque control on the EasyPad
4.5.10.1.1
Switching integrated torque control on/off
Switching torque control on
Torque control is assigned to the burr drive removed from the holder.
✔
The desired electric motor for torque control is removed from the
holder.
✔
The motor speed is shown on the EasyPad display.
➢ Press the
Endodontics / Purge
key.
If the integrated torque control is switched on, the status display
torque control
lights up.
The preset torque of 1.00 Ncm (Newton centimeters) is displayed
on the EasyPad display.
Switching torque control off
If the integrated torque control is switched on, when the burr drive
assigned to the torque control is removed from the holder, the torque or
speed of the contra-angle handpiece is displayed instead of that of the
motor on the EasyPad display. To show the motor speed again when the
burr drive is removed, the integrated torque control must first be switched
off.
➢ Press the
Endodontics / Purge
key briefly (< 1 s).
The
torque control
status display goes out. When the burr drive
is removed from the holder, the speed of the motor is shown on
the EasyPad display.
Purge
B
Purge