5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
Error message: Inserting instruments in the supply module
If an instrument is accidentally deposited in its holder during the reaction
time, the message
Insert instruments into water unit
is displayed on the
touchscreen after the simulation unit is switched on. The sanitizing
process is then completed anyway.
1. Remove the accidentally removed instrument and place it back in the
adapter on the supply module so that it can be rinsed in sanitation
phase 5.
2. If you want to exclude the instruments from sanitation, touch the
Exclude instrument
key.
Excluded instruments will not be rinsed in sanitation phase 5.
Sanitation phase 5 – Rinsing water lines with water
The DENTOSEPT P disinfectant is rinsed out of the SPRAYVIT hoses
and instrument hoses with water. This takes several minutes.
Sanitation phase 6 – Returning instruments to their holders
After the water lines have been rinsed, the instruments can be removed
from the cleaning adapters on the supply module and returned to their
holders.
✔
The
Return instruments to their holders
display appears on the
touchscreen.
➢ Place the SPRAYVIT sleeves and the straight and contra-angle
pieces on the instruments. Place all the instruments back in their
holders.
The sanitation process is finished. The simulation unit is ready for
operation again.
Displays in case of incomplete sanitation
The sanitation program detects whether the sanitation of all instruments
was completed. Errors that occurred during sanitation will be displayed on
the touchscreen as follows:
● Crossed-out, solid gray circle: The instrument was excluded from
sanitation before it had been filled with DENTOSEPT P disinfectant
● Orange circle with warning triangle: The instrument was not
(sufficiently) rinsed; the water lines still contain disinfectant
➢ In the latter case, rinse the affected instruments and tumbler filling
manually after sanitation.