5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
5.10.2
Sanitizing with the fresh water bottle
In the simulation unit, the purging and sanitizing functions and the fresh
water bottle are available as options. Both options must be present to
sanitize the water lines.
If the simulation unit is equipped with the purging and sanitizing option,
but does not have a fresh water bottle, the sanitizing function can be
selected on the EasyPad, however it has no function. Without the fresh
water bottle, only the purge function can be used.
If the simulation unit is equipped with a fresh water bottle, manual
sanitation must be performed:
● Regularly every 4 weeks
● After longer periods of disuse (> one week)
● If the germination index exceeds 100 germs per milliliter, see
“Microbiological water test”.
Sanitation takes at least 24 hours.
Additional devices, sanitation with disinfectant
Preparation
The following preparations should be made prior to beginning the
sanitizing process.
1. Set all burr drives and the SIROSONIC L scaler to the maximum
water flow rate. The water regulators under the instrument holder
must be turned counterclockwise.
2. Put all instruments and suction hoses in place.
NOTICE
Additional devices connected to the external device connection must
not be sanitized with the simulation unit.
Additional devices can be damaged. Residues of DENTOSEPT P
disinfectant can remain in the additional devices.
➢ Disconnect any additional devices from the simulation unit prior to
sanitation.
IMPORTANT
The water flow rate to the instruments is not checked by the simulation
unit.
➢ Make sure that the maximum water flow rate is set for all
instruments to be purged.