65 64 020 D3650
D3650
.
201.01.02
.
02 04.2016
103
Sirona Dental Systems GmbH
4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.6 Assistant element
båÖäáëÜ
Status control lamp
The status control lamp (D) signals the following states:
The mini LED has three operating modes, which you can select with the
Mode
H key:
Quick curing mode
In the quick curing mode, the Mini LED operates at full power for 10
seconds.
In this mode, the light intensity output is as follows:
● 1250 mW/cm
²
(± 10%) with the standard light guide, dia. 7.5 mm
● 2000 mW/cm
²
(± 10%) with the booster light guide, dia. 5.5 mm
Pulsed curing mode
In the pulsed curing mode, the Mini LED radiates in 10 consecutive light
pulses of 1 s each. There is a 250 ms break between the individual
pulses.
"Soft start" mode
The “soft start” mode features:
● A 10-second “soft start” from 0 to 1250 mW/cm
²
, or from 0 to 2000
mW/cm
²
with the "booster light guide," dia. 5.5 mm.
● Full power for 10 seconds.
Status control lamp
Mode
Off
Instrument in holder
green
Normalbetrieb
red, flashing
Overheat protection
0
5
10
Sec.
0
12.5
Sec.
0
5
10
15
20
Sec.