5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.10 Sanitizing
båÖäáëÜ
2. Remove a single instrument and hold it over the bucket. For the CS
dentist element, move the swivel arm of the instrument into the
working position.
3. Step on the pedal of the foot control briefly.
The removed instrument is then purged for approx. 15 seconds
with DENTOSEPT P disinfectant. Purging stops automatically.
4. Afterward, return the instrument to the instrument holder.
5. Repeat the procedure for all other instruments.
Tumbler filling, discolorations due to DENTOSEPT P
Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
➢ Switch the simulation unit off using the main switch on the supply
module. Make sure that the simulation unit is switched off for at least
24 hours, but not longer than 3 days (maximum sanitation time).
➢ Switch the simulation unit on again after 24 hours and before 3 days
have elapsed.
Filling the fresh water bottle with water
✔
The display
Refill water
appears on the EasyPad.
1. Unscrew the fresh water bottle from the holder on the support tube
and empty the fresh water bottle.
2. Mix food-grade water (< 100 CFU/ml) with the DENTOSEPT P
disinfectant at a ratio of 100:1 (1 liter of water, 10 ml DENTOSEPT P)
and fill this into the fresh water bottle. The bottle has a capacity of
approx. 1.3 liters.
3. Attach the hose to the fresh water bottle and screw the bottle into the
holder.
4. Press the favorite key
2 / Setup
.
A revolving element is displayed on the EasyPad next to the text
“AqU.”
max.
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.
2
Setup