5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.3 Instruments and instrument hoses
Operating Instructions Simulation Unit SIM MODULAR
Inserting water-carrying instruments into the cleaning adapters of the
supply module
The supply module has integrated adapters for water-carrying
instruments. These enable all instruments to be purged with water. For
this purpose, water-carrying instruments must be inserted into adapters
on the supply module.
1. Remove the SPRAYVIT sleeves from the valve bodies and the
straight and contra-angle handpieces from the water-carrying
instruments.
2. If the adapters are not yet located in the receptacles of the supply
module, insert them until they lock into place. The SPRAYVIT
adapters can be inserted only into the two receptacles on the right
side of the dentist element with the guide rib facing upward. Similarly,
a SPRAYVIT adapter must be inserted into the assistant element
side. The adapters remain in the supply module permanently.
3. Dentist element side: Insert the couplings of all water-bearing
treatment instruments into the adapter in the supply module (for
SPRAYVIT E: valve lever in the up position, locking knob in the down
position).
4. In addition for the CS dentist element: Move the swivel arm of the
instruments to be purged into the operating position so that the weight
of the instrument hoses allows them to remain in this position.
IMPORTANT
Arrangement of adapters
The adapters for the instrument couplings are color coded:
Yellow = SPRAYVIT E, water on right button
Orange = SPRAYVIT E, water on left button
White = highspeed handpiece
Blue = BL ISO E motor (ISO interface)
Red = SIROSONIC L scaler