02 04.2016
5 Care, cleaning and maintenance by the user
Sirona Dental Systems GmbH
5.10 Sanitizing
Operating Instructions Simulation Unit SIM MODULAR
Sanitation phase 3 – Treating water lines with DENTOSEPT P
disinfectant
The water is pumped out of the water tank of the supply module. Then the
water tank is automatically filled with pure DENTOSEPT P. This activates
flushing of the instrument hoses, the SPRAYVIT hoses and the tumbler
filler with the DENTOSEPT P disinfectant.
Tumbler filling, discolorations due to DENTOSEPT P
Sanitation phase 4 – Allow 24 hours reaction time
In order to effectively combat germs, you must allow the DENTOSEPT P
disinfectant to act at least 24 hours, but for no longer than 3 days
(maximum sanitation time).
✔
Sanitization phase 4 is highlighted on the touchscreen.
✔
The simulation unit displays the remaining reaction time on the
touchscreen next to the sanitation phase 4 field, starting from 24
hours.
✔
The simulation unit has automatically switched to Standby mode.
1. Leave all instruments plugged into the supply module.
2. Switch the simulation unit off using the main switch on the supply
module. Make sure that the simulation unit is switched off for at least
24 hours, but not longer than 3 days (maximum sanitation time).
3. Switch the simulation unit on again after 24 hours and before 3 days
have elapsed.
Stop
1
2
3
4
5
6
H2O
Desinf
NOTICE
Splashes of pure DENTOSEPT P can result in discoloration of surfaces
if left for too long.
Any splashes should therefore be removed immediately using a moist
cloth.
Stop
1
2
3
4
5
6
H2O
Desinf
23:59
IMPORTANT
Blocking water and air supply
If the simulation unit is turned on again after 24 hours, the sanitation
process continues automatically with sanitation phase 5. However, if the
water and air supplies are switched off, the DENTOSEPT P disinfectant
cannot be rinsed out of the water lines.
Switch the simulation unit back on after 24 hours only when the water
and air supply lines are open.