Sirona Dental Systems GmbH
4 Operation
Operating Instructions Simulation Unit SIM MODULAR
4.5 Dentist element
båÖäáëÜ
Calibrating the burr drive
Calibration is required when the integrated torque control is started, every
time the contra-angle handpiece is changed, and every time the contra-
angle handpiece is lubricated. Recalibration is not necessary when a file
is changed.
The contra-angle handpiece is automatically checked during calibration.
This includes a measurement of motor current at different speeds to
assess the properties of the system.
Calibration only with Sirona instruments
✔
The
Endodontics dialog
or the
Endodontics
sub-dialog is displayed
on the touchscreen.
1. Attach the contra-angle handpiece that you wish to use with torque
control to the electric motor.
2. Insert a file in the contra-angle handpiece. This ensures that the file
is taken into account in the measurement.
3. Touch the
Cal
key on the touchscreen.
or
➢
Press the right button of the C+ electronic foot switch.
The key flashes. The burr drive is ready for calibration.
4. Hold down the foot pedal throughout the duration of the calibration.
The
Cal
key continues to flash. During calibration, increasing
speeds are set on the motor. If the burr drive is calibrated, the key
permanently remains highlighted orange. Calibration is then
completed.
CAUTION
Use only Sirona instruments to guarantee correct calibration.
rpm
300
Ncm
1.00
Endo 3
Cal
2
1
0
S
rpm
300
Ncm
1.00
Endo 3
Cal
Cal
S
0
0
S
0