5 Care, cleaning and maintenance by the user
Operating Instructions Simulation Unit SIM MODULAR
5.3 Instruments and instrument hoses
båÖäáëÜ
All water-carrying instruments are inserted into the cleaning
adapters.
6. Press the favorite key
2 / Setup
.
Automatic purging of the water lines will start. A revolving
element is shown next to the
AutoPurge
display on the EasyPad.
Purging water lines
The removed instruments are purged with water for the duration of the set
purge time.
Following the purging process, the
End
display appears.
You can now continue in one of two ways:
● Leave the instruments in the supply module
The autopurge function remains activated.
The instruments remain in the supply module and the simulation unit
can be switched off at the mains. The next day, the autopurge
function is automatically performed again on all of the instruments
remaining in the supply module immediately after the unit is switched
on.
After, you can prepare the simulation unit for the next use.
If the simulation unit is out of operation over a prolonged period of
time, you can briefly switch it on every day and then switch it off again
when the purging process is finished. This ensures that the number
of microorganisms in the water lines will not increase excessively.
If any instruments are removed from the supply module or new
instruments are connected to the simulation unit while the latter is
switched off, you must deposit them in the adapters or in the
instrument holder again before switching the simulation unit back on!
● Deposit instruments
The autopurge function is completed.
➢ Remove the instruments from the supply module and put them back
in place.
The autopurge procedure is finished. The simulation unit is ready for
operation again and can be prepared for the next use.
IMPORTANT
Pinching the instrument hoses
Be careful not to pinch the instrument hoses when inserting the
instruments.
If the hoses are pinched, the water flow will be obstructed during
purging.
The water flow rate to the instruments is not checked by the simulation
unit.
2
Setup