Sirona Dental Systems GmbH
13 List of messages
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
13.2 Error message structure
59 87 594 D 3352
D 3352.201.01.18.02
115
13.2
Error message structure
The error messages are displayed in the form of an error code. They are not provided in plain-text form.
13.2.1
The error message code has the following structure:
All error messages of the system are grouped according to these criteria.
13.2.2 Ex
Digit (x) is intended to provide the user with quick help in deciding how to deal with this error.
Ex yy zz
Ex
Error type/"troubleshooting" classification for the user
yy
Location; module; subsystem or logical function unit
zz
Consecutive number with error ID
Ex
Description
Actions required
Error group
1
System warning;
system message
Acknowledge the error message. Contact
your Customer Service. Continued
operation of the system is ensured.
This error group includes all errors that
indicate still acceptable tolerance
variations, or messages about states
which do not directly affect system
operation.
2
Errors caused by
system overload
Acknowledge the error message. Repeat
the procedure step after a certain waiting
time. If the error message reappears,
prolong the waiting time. If the error state
persists, contact your Customer Service.
This error group includes states that
indicate, for example, temporary excess
temperatures or the like. The cause of the
error disappears automatically after a
certain waiting time.
3
The system detects
that a key was pressed
during power-on.
Switch the system off and back on; if the
error reoccurs without any user
intervention, contact your Customer
Service.
CAUTION!
After switching the unit off with
the main switch, you must wait for
approx. 2 minutes before switching it back
on.
This error group includes all errors that
indicate invalid signal states of keys and
safety signals during power-on.
4
Malfunction or
mechanical obstruction
of unit movements
Acknowledge the error message; make
sure that the movements of the unit are not
obstructed. Repeat the last procedure step
or exposure. If the error reoccurs without
any identifiable cause -> contact your
Customer Service
This error group includes all errors that
indicate problems with the motor-controlled
movements on the outside of the unit.