11 Transversal slices (TSA)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
11.2 Patient positioning
59 87 594 D 3352
D 3352.201.01.18.02
97
In cases where you need more information in the central
facial area you can use the
green
bite block (or contact
segment) included in the scope of supply.
This positions the patient a little lower in relation to the
beam path.
The light localizer beam
(A)
, which reflects also on the
sensor, can be used as an auxiliary tool. Set the light
beam to the upper edge of the TSA display on the sen-
sor.
The simultaneous reflection of the light beam
(A)
on the
patient’s face marks the top edge of the exposure and
helps you decide whether you need the
blue
or the
green
bite block (or contact segment).
ATTENTION
If you have a patient without anterior teeth, always place
a cotton pellet between his/her jaws as illustrated!
A