10 Cephalometric exposures (CEPH)
Sirona Dental Systems GmbH
10.7 Releasing a cephalometric exposure
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
86
D 3352.201.01.18.02
NOTE
i
Please note that any program settings which may have
been changed must be preselected again before repeat-
ing the exposure.
Automatic exposure blocking
(thermal protection of the tube)
Premature release of a new exposure is prevented by
the automatic exposure blocking function.
When you press the exposure release button, the mes-
sage "
Ready for exposure in "XX" seconds
" appears in the
comment line of the touchscreen.
The remaining cooling time is counted down and is dis-
played under
"XX"
.
Only after the cooling period has elapsed is it possible to
release a new exposure.
NOTE
i
If the exposure sequence P1–C3–C4 (in this order) has
been selected in SIDEXIS, there are shorter cooling pe-
riods between the individual programs.
However, after the entire program sequence has been
processed, a longer cooling period is required.