10 Cephalometric exposures (CEPH)
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
10.3 Positioning a ceph patient
59 87 594 D 3352
D 3352.201.01.18.02
79
C1 p.a. views with shadowing in the thyroid
area
(only for “half-axial skull radiograph”, off-axial cra-
nial overview)
This exposure technique with the head reclined and the
mouth open fully also supports shadowing in the thyroid
area.
To do this, touch the head icon in the Submenu column
(8), and select the shadowing icon in the submenu line.
It will then be displayed in the column (8).
Shadowing results in a dose reduction in the thyroid
area.
PAN
9,1s
80kV
14mA
?
CEPH
C1 p.a.
C1 p.a.
1558
9,1s
TS
Ready for exposure
60
90
120
70
80
100 110
C1
p.a.
8