10 Cephalometric exposures (CEPH)
Sirona Dental Systems GmbH
10.2 Preparations on the cephalometer
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
74
D 3352.201.01.18.02
10.2
Preparations on the cephalometer
Inserting the sensor
If you operate the unit with one sensor only, you must
remove the sensor (9) from its slot in the rotating ele-
ment for pan exposures and plug it into the slot on the
cephalometer.
NOTE
i
To remove the sensor, hold it firmly, press the pushbutton
(8) fully in and hold it down. Remove the sensor from its
holder by pulling it downward.
ATTENTION
DO NOT DROP THE SENSOR!
A shock sensor for detecting shocks or drops is built in.
ATTENTION
When removing the sensor from the pan/ceph slot or
handling a sensor that has already been removed, make
sure not to touch the sensor plug on the unit end, espe-
cially not while touching the patient at the same time.
Slide the sensor with its two guide bolts into the guide
sleeves and push it up to the stop; it is not necessary to
press the pushbutton (8) when doing this.
If the sensor (9) is not fully inserted into the slot on the
cephalometer, the help message
"Plug sensor into
Ceph slot"
appears in the comment line.
If a PAN sensor has been plugged into the Ceph slot
unintentionally, the help message
"Plug in Ceph sen-
sor"
is displayed.
NOTE
i
If you operate the unit with two sensors, the PAN sensor
may remain in its slot on the rotating element.
•
Seize the nose support (18) at its upper front end,
pull it toward the front as far as possible and fold it up
toward the side.
•
Seize the ear plug holders (25) at their upper ends
and push them outward as far as possible.
NOTE
i
All of the following illustrations of the cephalometer are
shown in the left-handed arm version. However, they al-
so generally apply to the right-handed version of the ce-
phalometer, see adjacent illustration.
T
R
CEPH
CEPH
PAN
PAN
TS
TS
LS
LS
Filmkassette einrasten
1260
64
62kV
8mA
8mA
?
P6.1
12,1s
CEPH
CEPH
PAN
PAN
TS
Filmkassette
einlegen
Filmkassette
einlegen
1260
68kV
P6
12,8s
AEC
8mA
8mA
20
30
40
50
20
30
40
50
9
8
9
18
25
Left-handed arm
Right-handed arm
Содержание Orthophos XG Plus DS/Ceph
Страница 1: ...b loqelmelp udm apL l f...
Страница 127: ......