9 Operation
Sirona Dental Systems GmbH
9.3 Positioning the patient
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
59 87 594 D 3352
64
D 3352.201.01.18.02
Temporomandibular joint views, programs
TM3 – TM6 with temporomandibular joint
supports
•
For temporomandibular joint views, you must fit the
temporomandibular joint supports (C)
"1"
on the
right and (D)
"2"
on the left side in place of the tem-
ple supports (7).
To do so, remove the two temple supports (7) after
pressing the corresponding locking button; instead,
push in the two temporomandibular joint supports
(C) and (D) until they engage.
Two sterile ear holders (E) must be plugged into the
temporomandibular joint supports (C) and (D).
•
Remove the bite block or chin rest.
NOTE
i
For devices until October 2006:
Please observe the orientation (right/left) of the temporo-
mandibular joint supports specified on page 28.
•
Using the “
up
” or “
down
” arrow keys on the Easy-
pad, adjust the height of the unit so that the ear hold-
ers are at the height of the external auditory canals.
•
Position the patient’s head between the temporo-
mandibular joint supports (C) and (D).
Close the temporomandibular joint supports with
key (37) on the Easypad so that the ear holders fit in-
to the external auditory canals.
•
For programs
TM1
,
TM3
and
TM5
, the patient’s
head must be positioned in the
F
rankfort
H
orizontal
plane
FH
.
•
For programs
TM2
,
TM4
and
TM6
, the patient’s
head must be
inclined towards anterior
in relation
to the FH so as to largely prevent overlaps.
•
Align the center of the anterior teeth or of the face
with the
central light line
.
2
1
E
7
C
D
T
R
CEPH
PAN
TS
LS
Filmkassette einrasten
1260
64
62kV
8mA
?
P6.1
12,1s
CEPH
PAN
PAN
Aufnahmebereit
Aufnahmebereit
1260
64
64kV
8mA
?
P1
14,2 s
14,2 s
TS
1
37
C
D