9.3 Positioning the patient
59 87 594 D 3352
D 3352.201.01.18.02
63
Cleaning the occlusal bite block
If the hinges of the occlusal bite block begin to emit
squeaking noises in operation following longer periods of
use, they must be cleaned.
To do this, proceed as follows:
1.
Pull the occlusal bite block out of its holder on the
ORTHOPHOS XG
Plus
.
2.
Carefully push apart the guide mandrel of the lever
(1) on the bite block plate and the eyelet of the con-
necting rod (2) slightly in the directions of the arrows
and unhinge the lever.
3.
Swing the bite block (3) upward into a vertical posi-
tion so that the lever (1) is pointing downward.
4.
Pull the bite block plate (3) forward out of its hinge.
5.
Clean the hinge axles (A) and the guide lugs (B) with
disinfectant.
6.
Assemble the occlusal bite block by following the
same procedure in reverse order. When assembling,
check the position of the bite block plate; the seg-
ment must point toward the connecting lever.
Insert the occlusal bite block into its holder on the
ORTHOPHOS XG
Plus
.
1
2
3
1
3
A
A
A
B
B
B