Sirona Dental Systems GmbH
9 Operation
Operating Instructions ORTHOPHOS XG
Plus
DS/Ceph
9.8 Releasing the exposure
59 87 594 D 3352
D 3352.201.01.18.02
69
•
The forehead support and the temple supports open
automatically, and the patient may leave the unit.
After completion of the exposure
the X-ray image is displayed on the PC monitor in
SIDEXIS.
In addition, a small control image (A) is displayed on the
touchscreen; it is not suitable for diagnostic purposes.
You
must
close the preview image again by touching
the touchscreen.
Display of the dose area product
The exposure mode, exposure program, tube voltage,
tube current, real radiation time,
dose area product
and
shadowing (depending on the exposure program) are
again displayed on the touchscreen.
•
Acknowledge the exposure time actually needed by
pressing the return key
R
.
•
Then reset the rotating element to its starting posi-
tion by pressing the return key
R
a second time.
Canceling an exposure
If you let go of the exposure release button prematurely,
the exposure is canceled.
The exposure time and dose area product display read-
ings flash following exposure cancellation. The exposure
time which had elapsed prior to cancellation is dis-
played.
•
Press the
"R"
key on the Easypad twice.
NOTE
i
Please note that any program settings which may have
been changed must be preselected again before repeat-
ing the exposure.
•
After the rotating element has returned to its starting
position, repeat the exposure.
T
R
P1
R
A
PAN
P1
64kV
8mA
14,1s
46mGycm²
H320 - R button,
confirm exposure data
H320 - R button,
confirm exposure data